<In case of using washer disinfector (automatic washing)>
Temperature
Stage Procedure
[°C (℉)]
I
Preclean <25 (77)
II
Washing 55 (131)
III
Rinsing <25 (77)
Hot water
IV
disinfection
V
Drying
2) Inspection
Check clips visually one by one and do not use if the
following abnormalities are found.
・Damage
・Deformation
・Corruption
・  I nappropriate blade position (deviation, clearance,
tip unevenness, etc.)
・Remaining dirt
Time
Quality of
[min]
water
3
Tap water
10
Tap water
Deionized
4
water
Deionized
water
-17-
Alkaline detergent (such as cleaning
agent for aluminum) that does not
affect anodic oxidation coatings
Diluted concentration 0.5%
Set rinsing process program as
"1 time (1 minute) × 4 times = 4
minutes"
Follow the program of the device.
Remarks