When the order of magnitude is not known, place the switch on the highest range
then lower progressively to the appropriate range.
Reading:
To obtain the value of the current in A, multiply the reading on the appropriate scale
by the reading coefficient, as indicated below.
Warning: Limit the measurements time to 3 minutes above 2A AC.
A ranges
50µA
1mA
DC
10mA
100mA
1A
150µA
0,3mA
3mA
AC
30mA
300mA
3A
(1)
Position common to the 100mV Dc range
(2)
Position common to the 3V AC range.
30
Scales
(1)
100 black
100 black
100 black
100 black
100 black
(2)
30 black
30 red
30 red
30 red
30 red
30 red
Reading
coefficient
X 0.5
X 0.01
X 0.1
X 1
X 0.01
X 5
X 0.01
X 0.1
X 1
X 10
X 0.1
Protection
160mA fuse
3.15A HPC fuse
Idem 3V~
160mA fuse
3.15A HPC fuse
C.A 406