6.6 ORP PROBE MENU

• The next screen shown asks if the ORP probe is currently connected and to be used (refer Figure
6.19). Unlike the pH probe, the ORP probe needs to be rinsed with tap water and then left to soak in its
calibration solution for 10 minutes prior to calibrating. You may wish to elect to come back this step later,
or if you've chosen not to use the ORP probe, skip to step 6.7 of this manual. The ORP probe can always
be re-calibrated after initial installation (refer Section 8.2.2);
Figure 6.19
• If you've chosen not to use the ORP probe, skip to step 6.7 of this manual;
• Use the
menu up/down buttons to toggle between yes and no. Once correct, then
press
menu/setting select;
• If you've chosen to use the ORP probe, the next screen will instruct you to put the ORP probe into the
ORP solution (ref Figure 6.20);
Figure 6.20
• Carefully unscrew the bottom half of the bottle, from the bottle's lid. Follow the same process outlined
for the pH probe (refer Section 6.5).
• Once the ORP probe is bathing in its solution, press
• The screen will next display ORP calibration in progress (refer Figure 6.21). A count will commence.
Typically, the calibration will take less than 15 seconds, but may take up to a minute;
is the orp probe
connected?
->yes
ORP calibration:
put orp probe
into solution
->OK
menu/setting select;
24