Télécharger Imprimer la page

Zeiss Axio Observer Mode D'emploi Succinct page 11

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
Axio Observer
Line voltage of external power supply unit of Axio Observer 7 materials ........... 100 V to 240 VAC ±10 %
Mains frequency ................................................................................................................ 50 Hz – 60 Hz
Power consumption Axio Observer 3, 3 materials and 5, 5 materials ..................................... max. 300 VA
Power consumption of external power supply unit of Axio Observer 7, 7 materials ............... max. 190 VA
Power supply unit for HBO 100
Operating area ................................................................................................................Enclosed rooms
Protection class ..................................................................................................................................... I
Ingress protection rating ................................................................................................................. IP 20
Line voltage ................................................................................................................ 100 V to 240 VAC
Mains frequency ................................................................................................................ 50 Hz – 60 Hz
Power consumption if operating with HBO 100 ............................................................................ 155 VA
Fuses in accordance with IEC 127
Microscope stand Axio Observer 3, 3 materials and 5, 5 materials ................... T 5 A/H / 250 V, 5x20 mm
Power supply unit VP232-2 for Axio Observer 7, 7 materials .......................... T 4.0 A/H / 250 V, 5x20 mm
HBO 100 power supply unit (ballast unit) ...................................................... T 2.0 A/H / 250 V, 5x20 mm
Light sources
HBO 50W/AC mercury vapor short-arc lamp
Output ....................................................................................................................................... 50 W
Average service life ................................................................................................................... 100 h
HBO 103 W/2 mercury vapor short-arc lamp .................................................................................. 100 W
Desktop power supply unit for external filter wheel
Protection class ..................................................................................................................................... I
Desktop power supply unit input ........................................................ 100 V to 240 V ±10%, 50 – 60 Hz
Desktop power supply unit output ............................................................................ 24 VDC, max. 2.5 A
Optical-mechanical data
Stand with stage focusing ............................................ with coarse focusing drive approx. 2 mm/rotation
Change of objective ......................................................................... via 6-position nosepiece, H DIC M27
Objectives ........................................................................................................... with M27 screw thread
Eyepieces ..................................................................................Plug-in diameter 30 mm, field number 23
09/2017
TECHNICAL DATA
and fine focusing drive approx. 1/10 coarse/fine focus transmission ratio,
431004-6044-008
A change of the line voltage is not required!
total travel approx. 10 mm, 13 mm also possible
ZEISS
9

Publicité

loading

Ce manuel est également adapté pour:

Axio observer 3Axio observer 5Axio observer 7