The TwisTorr Medium-TMP onboard Controller belongs to the second
CAUTION!
installation (or overvoltage) category of the EN 61010-1 standard. Therefore,
connect the device to a mains line that complies with the above category. The
TwisTorr Medium-TMP onboard Controller has Input/Output and serial
communication connectors that must be connected to external circuits in such a
way that no live parts of electronic board are accessible. Make sure that the
insulation of the device connected to the TwisTorr Medium-TMP onboard
Controller is adequate even in the case of a single fault as per the EN 61010-1
standard.
For installation of optional accessories, see "Technical Information".

Use

This paragraph details the main operating procedures. Prior to
operating the Turbo controller, make all vacuum manifold and
electrical connections, referring to the Turbo pump instruction
manual.
TwisTorr Medium-TMP onboard Controller User Manual / 87-901-050-01 (B)
Instructions for Use