17
Technical Information
Procedure to Connect the Serial and I/O Ports to an External
Cable
Window-Meanings
Window-Meanings
N.
000
001
008
100
101
102
103
104
105
106
164/181
TwisTorr Medium-TMP onboard Controller User Manual / 87-901-050-01 (B)
Read/Write
Datatype
R/W
L
R/W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
R/W
N
R/W
L
Description
Start/Stop (in remote mode
the window is a read only)
Low Speed (in remote mode
the window is read only)
Remote (default) or Serial
configuration
Soft Start
(write only in Stop condition)
Set Point type
Set Point threshold (expressed
in Hz, mA or s)
Set Point delay: time
between the pump start and
the set point check (seconds)
Set Point signal activation type:
the signal can be "high level
active" or "low level active"
Set point hysteresis
(in % of threshold)
Cooling agent
Admitted Values
Start = 1
Stop = 0
0 = OFF
1 = ON
(default = 0)
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 643)
0 to 99999
(default = 0)
0 = high level
active
1 = low level
active
(default = 0)
0 to 100
(default = 2)
0 = AIR
1 = WATER
(default = 0)