TwisTorr Medium-TMP onboard Controller User Manual / 87-901-050-01 (B)
Procedure to Connect the Serial and I/O Ports to an External Cable
R/W
N
W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
Reserved to Agilent service
R/W
L
R/W
N
Reserved to Agilent service
Technical Information
Baud rate
Pump life/ cycle time/ cycle
number reset
Interlock type (default = 1)
Analog output type: output
voltage signal proportional to
frequency or power
Low speed setting [Hz]
Rotational frequency setting
(Hz)
Maximum rotational
frequency in Hz (active only
in Stop condition)
Set vent valve on/off
(on = closed)
Set the vent valve operation
Vent valve opening delay
(expressed in 0.2 sec)
17
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write
'1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
680 to WIN 121
(default = 680)
680 to WIN 121
(default = 825
for X3512-64016,
714 for X3512-
64006 )
680 to 825
(default = 825)
On = 1
Off = 0
(default = 1)
Automatic = 0
(see note 1.)
On command =
1 (see note 2.)
0 to 65535
(corresponding
to 0 to 13107
sec)
165/181