Signal characteristics and use
Pin N.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Signal Description
Start/Stop: input signal to start or stop the pump. With the supplied
cover connector the START/STOP (+) signal is connected to the +24
Vdc pin and the START/STOP (-) signal to the GROUND pin: in this
condition the pump automatically starts as soon as the controller
recognises the input supply ("Plug & Pump").
Interlock: input signal to control the pump rotation. With the
supplied cover connector the interlock (+) signal is connected to the
+24 Vdc pin and the interlock (-) signal to the GROUND pin.
SOFT START: this input is used to provide a "soft start" to the pump;
see chapter SOFT START later on in this document.
TwisTorr Medium-TMP onboard Controller User Manual / 87-901-050-01 (B)
Signal name
Start/Stop (+)
Start/Stop (-)
Interlock (+)
Interlock (-)
Speed setting (+)
Speed setting (-)
Soft start (+)
Soft start (-)
+ 24 vdc
Contact: normally open
Programmable set point (+)
Contact: normally open
Fault output
Programmable analog signal (+)
• Ground
• Programmable analog signal (-)
Technical Information
Interconnections
In-/Output
In
In
In
In
In
In
In
In
Out
Out
Out
Out
Out
149/181
17