Télécharger Imprimer la page

Parameters; Auto Power Off - Adam Equipment CORE Série Mode D'emploi

Masquer les pouces Voir aussi pour CORE Série:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EN 
4.0  PARAMETERS
The balance has 3 parameters that can be set by the user. 
FUNCTION 
Auto Off 
Backlight 
Weighing unit 
 
4.1  AUTO
POWER
 
The auto power off function helps conserve power when using the batteries or
The Auto switch-off time may be set up by the user and is the time after a period of
inactivity i.e. no weight movement or key presses.
1) To set this parameter turn the power off, press and hold the [Tare] key then 
press the [on/off] key once, then release the [Tare] key. 
2) The display will show the first function XX PWR, where XX is the last setting.  
3) Press [Tare] to select the settings as follows: 
Off PWR
10 PWR
20 PWR
30 PWR
40 PWR
50 PWR
60 PWR
70 PWR
80 PWR
90 PWR
 
4) Press the [Unit] key to confirm the setting 
5) To exit turn the balance off and back on again 
© Adam Equipment 2016
 
 
SECTION 
See section 5.1 
See section 5.2 
See section 5.3 
OFF 
 
Auto power off
10 seconds before the power will switch off
20 seconds before the power will switch off
30 seconds before the power will switch off
40 seconds before the power will switch off
50 seconds before the power will switch off
60 seconds before the power will switch off
70 seconds before the power will switch off
80 seconds before the power will switch off
90 seconds before the power will switch off
DESCRIPTION
Sets the auto power off function 
Sets the backlight 
OFF BL backlight is off 
1 BL backlight is automatic 
2 BL backlight is always on
Sets the units to be used
kg – g –  lb – oz – ct – dr – GN – N‐tl.t 
10
 
308669539revB6-Mar17
DC
adapter.

Publicité

loading