6.3

Sleep Mode

Sleep Mode
Sleep Mode
6.4

Additional information on the unit operations

NOTE
If the hand piece is placed in the charging unit before the curing cycle is terminated, the light emission
automatically switches o .
When the hand piece is placed in the charging unit its functioning is inactive. The operativeness of the
" Start / Stop " button
The Translux Wave is equipped with microprocessors which continuously check the optimum charging
parameters of the battery. After every treatment, the hand piece should therefore be inserted in the charging
unit for storage irrespective of the battery level.
7

Alarm conditions

7.1

Faulty battery

beep
IMPORTANT
This error condition is detected only when the hand piece is placed in the charging unit.
7.2

Faulty LED

7.3

Overheat protection

Start / Stop
NOTE
The hand piece becomes operational upon reaching operational temperature. No " manual reset " required.
7.4

BUILT-IN Temperature overload sensor defective

NOTE
If the temperature overload sensor fault occurs during an exposure cycle, the software allows
completing the exposure. No further exposure cycle is allowed by pressing the " Start / Stop " button
" 4 beeps " are emitted.
Sleep Mode
Start / Stop
is prevented.
Fig. a
10s
4 beeps
Fig. b
3 beeps
Sleep Mode
blinking