7.5
Battery-low signal
8

Cleaning, disinfection and sterilisation

DISCONNECT FROM MAINS
Before carrying out the cleaning and disinfection procedures disconnect the charging base from the electrical
power source!
WARNING
Control of infections
GB
The light guide and the protective light cone are delivered in NON STERILE condition and must be cleaned,
disinfected and sterilised (steam) before being used for the first time as well as before every use.
Do not sterilise the hand piece and the charging unit!
- 16 -
Do not sterilise the battery!
8.1
Disassembly
8.2
LIGHT GUIDE – Cleaning and disinfection
CAUTION
Do not use sharp or pointed instruments to clean the light guide, since they could scratch the surface of the
light guide and thus decrease light transmission.
Do not use cleaning / disinfectant agents containing organic, mineral and oxidising acids, hydrogen peroxide,
chloride, iodine, bromide and solvents.
1)
2)
8.2.1
LIGHT GUIDE – Inspection before the sterilisation
WARNING
Do not use a damaged light guide. If damage is apparent, replace the light guide with a new one.
If the light guide is still contaminated, repeat the cleaning and disinfection procedure.
8.3
PROTECTIVE LIGHT CONE – Cleaning and disinfection
CAUTION
Do not use cleaning / disinfectant agents containing organic, mineral and oxidising acids, hydrogen peroxide,
chloride, iodine, bromide and solvents.
2 beeps
Charge the battery!
2 beeps
Charge the battery!
Allow the disinfectant solution to dry.
Allow the disinfectant solution to dry.