Exposure 10 seconds polymerisation
Acoustical signal:
The following acoustical signals (beep) are emitted:
1 beep
1 beep
Exposure 20 seconds "Soft-Start" polymerisation
Acoustical signal:
GB
The following acoustical signals (beep) are emitted:
1 beep
- 14 -
1 beep
1 beep
1 beep
NOTE
Acoustical signal:
6.2

Measurement of the Light Intensity

CAUTION
Before each treatment always check the light intensity by using the light intensity meter
integrated in the Translux Wave charging unit.
• Without applying pressure, place the light guide flat and directly on the testing area ( 8
GREEN LED
YELLOW LED
CAUTION
If the light intensity is insu cient, please carry out the following controls and measures:
Use original Heraeus Kulzer light guides only.
WARNING
If the measures described above do not result in any improvement, do not use the device for any treatment.
Switch o the unit (disconnect the charging unit from the electrical power source) and ensure that
unauthorised unit re-activation is not possible. Contact a Heraeus Kulzer's service partner, or technically
trained personnel authorised by Heraeus Kulzer (see paragraph 14 Service).