Push the defective fuse out of the fuse holder completely.
-
Insert a new fuse which has the same rated current, same rated voltage, same breaking capacity, same
triggering characteristics and same dimensions.
-
Lock the battery compartment cover into place on the bottom part and tighten the screws.
See figure 8:
Fuse replacement
9.5 Calibration
Benning guarantees compliance with the technical and accuracy specifications stated in the operating
manual for the first 12 months after the delivery date.
To maintain the specified accuracy of the measurement results, the instrument must be recalibrated at
regular intervals by our factory service. We recommend a recalibration interval of one year. Send the unit
to the following address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt

9.6 Spare parts

Fuse 1.6 A, 1000 V, superfast-acting (FF), breaking capacity ≥ 30 kA, D = 6.3 mm, L = 32 mm part no.
10194027
10. Environmental notice
At the end of the product's useful life, please dispose of it at appropriate collection points provided
in your country.