Error limits

Error limits according DIN EN ISO 8655-5 related to the nominal capacity
(= maximum volume) indicated on the instrument, obtained when instrument and
distilled water are equilibrated at ambient temperature (20 °C/68 °F). Testing
takes place according DIN EN ISO 8655-6 with a completely filled instrument
and with uniform and smooth dispensing.
Error limits Dispensette
Nominal volume
ml
10
* A = Accuracy, CV = Coefficient of Variation
Partial volume
The percentage values for A and CV are relative to the nominal volume (V
and must be converted for partial volumes (V
V
N
A
=
· A
V
T
N
T
e.g.
volume
V
N
V
= 50% N
T
V
= 10% N
T
* A = Accuracy, CV = Coefficient of Variation
Note:
The error limits are well within the limits of DIN EN ISO 8655-5. The maxi-
mum error limit for a single measurement can be calculated EL = A + 2 CV
(for volume 10 ml: 50 µl + 2 x 10 µl = 70 µl).