Tecno-gaz ONYX B 5.0 Mode D'emploi page 82

Autoclave
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89
ENGLISH
15
Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
If an alarm message is diplayed (code "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors hold down at the same time the buttons under the bar indicated
with Reset.
ERROR
A 001
Cycle interrupted by the user
A 101
Vacuum not archieved in 10 min.
Vacuum not maintained on the first phase
A 111
of VACUUM TEST
Vacuum not maintained on the second
A 121
phase of VACUUM TEST
During preheating phase the autoclave did
A131
not load right water amount
A132
Error in the flow-meter operation
Pressure above the allowable limit during
A133
the water recall for the warmup phases
A 200
Error in the control of the EV operation
A 400
A 401
Error in the locking operation
A 403
A 405
A 551
Pressure out of limit
A 637
Error in the access to the memory card
Reading of the T1 probe during sterilization
A 651
above the maximum limit
38
MESSAGES OF ERRORS
CAUSE
SOLUTION
• Reset and restart the system
• Check the gasket
• Check the door adjustment
• Reset and restart the system
• Check the gasket
• Check the door adjustment
• Reset and restart the system
• Check the gasket
• Check the door adjustment
• Reset and restart the system
• Reset and restart the system
• Reset and restart the system
• Reset and restart the system
• Reset and restart the system
• Reset and restart the system
• Let cool the room
• Reset and restart the system
• Check that the SD card is
correctly inserted
• Chech SD card: lever might
have moved on "lock" position
• Reset and restart the system
• Let cool the room
• Reset and restart the system

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Onyx b 8.0Ab015zxyAb017zxz

Table des Matières