Communication Failures
If your PDM is damaged or not working as
expected, call Customer Care for assistance. Be
sure to check your blood glucose frequently.
Remove your Pod and contact your healthcare
provider for treatment guidelines to follow after
you remove it.
3. Move across the room or to a different room. Then either:
a. Press Discard to begin the deactivation process.
b. Press Retry to repeat the attempts to communicate with
the Pod. The PDM tries to communicate with the Pod
again. If it cannot communicate, it displays the same
screen as in Figure 11-13, then finally a second "Bolus
status unknown" screen (Figure 11-14). Press OK to check
the Pod status or press Discard to begin the deactivation
process.
4. If at any point in this process the PDM can communicate again
with the Pod, you will see Figure 11-15. The message indicates
that your original instruction to cancel the bolus was
successful.
a. Press OK. The PDM displays the screen in Figure 11-16,
showing how much insulin was delivered before the bolus
was canceled.
b. Press OK again to return to the Status screen.
138
Figure 11-14
Figure 11-16
Figure 11-15