Enter Minimum Bg Allowed For Use In Bolus Calculation; Enter Insulin-To-Carbohydrate Ratio - OmniPod CAT45E Guide De L'utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. Use the Up/Down Controller buttons to enter the "correct
above" value or correction threshold (the BG value above
which you want the PDM to suggest a correction bolus), then
press Next.
If the suggested bolus calculator is On, a correction
bolus will be suggested only when your BG is above
the value shown in brackets on the screen.
3. To save the target BG value, press Done. To change these
set tings, select the segment you wish to change and press
Edit.
You can define up to 8 target BG time segments. To enter
additional segments, choose [add new], then:
a. Press New.
b. Enter a start time for the target BG segment, then press
Enter.
c. Enter an end time for the target BG segment, then press
Next.
d. Enter a target blood glucose value, then press Next.
Use the Up/Down Controller buttons to enter the
"cor rect above" value or "correction threshold" (the BG
value above which you want the PDM to suggest a
correction bolus), then press Next.
e. When you have finished entering segments, press Done.
Getting Started
4. To see the segments as a list, press List. To return to the
graphic view, press Graph.
5. Press Save to accept the target BG profile you have set.

Enter minimum BG allowed for use in bolus calculation

1. Use the Up/Down Controller buttons to enter a minimum
blood glucose value, below which the System will not
calcu late a suggested bolus. The values are in 0.1 mmol/L
incre ments and range from 2.8 to 3.9 mmol/L.
2. Press Next.

Enter insulin-to-carbohydrate ratio

1. Use the Up/Down Controller buttons to enter your insulin-
to-carbohydrate ratio, then press Next. (Insulin-to-carbohy-
drate ratio is the number of carbs covered by 1 unit of insulin.)
2. To save the ratio, press Done. To change the ratio, select the
segment you wish to change and press Edit.
You can define up to 8 insulin-to-carbohydrate time
segments. To enter additional ratios, choose [add new], then:
a. Press New.
b. Enter a start time for the segment, then press Next.
c. Enter an end time for the segment, then press Next.
d. Enter the insulin-to-carb ratio for the new time segment,
then press Next.
e. To save the new ratio, press Done.
2
13

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Cat45f

Table des Matières