Alert error message
Possible causes
Meter Error 3
• Incorrect test procedure.
For example, putting blood
on the test strip before
See user manual for
inserting the test strip into
further instructions.
the meter or applying
blood before the Blood
Drop and Test Strip symbols
display.
• Problem with the test strip.
• Problem with the meter.
Meter Error 4
• Problem with the test strip.
• Problem with the meter.
See user manual for
further instructions.
Meter Error 6
• Blood sample was applied
while the code number was
being changed.
See user manual for
further instructions.
Action
1. Be sure you see the Blood Drop and Test Strip symbols on the display
screen before you apply blood or control solution.
2. Conduct a control solution test using a new test strip. If the results of
the control solution test are within the range printed on the side of
the test strip vial, retest using blood and a new test strip.
3. If the control solution test does not work or the error persists, call
Customer Care.
1. Conduct a control solution test using a new test strip. If the results of
the control solution test are within the range printed on the side of
the test strip vial, retest using blood and a new test strip.
2. If the control solution test does not work or the error persists, call
Customer Care.
1. Retest using a new test strip. Do not press any buttons or Soft Keys
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care.
Alerts and Alarms
10
125