OPERATION MANUAL
CLASSIFICATIONS OF EQUIPMENT
• Type of protection against electric shock:
– Class II equipment
• Degree of protection against electric shock:
– Type BF applied part:
• Method of sterilization or disinfection recommended by Brasseler USA:
– See page 11. Sterilization
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529:
– Foot Control: IPX1 (Protected against vertically falling water drops)
• Degree of safety of application in the presence of a fl ammable anesthetic mixture with air or with oxygen or nitrous oxide:
– EQUIPMENT not suitable for use in the presence of a fl ammable anesthetic mixture with air or with oxygen or nitrous oxide.
• Mode of operation:
– Continuous operation
INTENDED USE
This product is intended only for dental clinic/dental offi ce use. This device generates ultrasonic waves
intended for use in dental applications such as scaling, root canal treatment, periodontal and cavity preparation.
RX ONLY CAUTION: U.S. Federal law restricts this device to sale by or on the order of a dentist.
CAUTIONS FOR HANDLING AND OPERATION
• Read these cautions carefully and use only as intended or instructed.
• Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device. Safety instructions
are classifi ed as follows in accordance with the seriousness of the risk.
CLASS
WARNING
CAUTION
NOTICE
* OPERATION PRINCIPLE
A sinusoidal electrical signal, at ultrasonic frequency ( f > 20kHz ) , is delivered by the generator. This signal is applied to the 'piezoelectric ceramic' located inside
the transducer. Piezoelectric ceramic converts this signal into mechanical vibrations. These vibrations are at the same ultrasonic frequency as the electrical
signal. The mechanical vibrations are propagated towards the distal end of the transducer. The "TIP" insert, which is attached at the distal end of the transducer,
vibrates at ultrasonic frequencies and makes it possible to achieve the aimed purpose.
DEGREE OF RISK
A hazard that could result in bodily injury or damage to the device if the safety instructions are not
followed.
A hazard that could result in light or moderate bodily injury or damage to the device if the safety
instructions are not followed.
General information needed to operate the device safely.