12. TROUBLESHOOTING

When problem occurs, please review the following list prior to contacting Brasseler USA.
PROBLEM
PROBABLE CAUSE
The Power Indicator
is not on, even
when the power is
ON.
No / Poor
The Tip does not
vibration.
generate vibration,
even when
depressing the Foot
Control.
The Tip is
bent or broken.
The Tip is
fl ying away.
Noise from
the handpiece.
The handpiece
is overheating.
The water does
not reach the
Control Unit.
No / Poor water.
Check to see if
water reaches the
Control Unit.
Water is leaking from
the joint between
Water leakage.
the handpiece and
the cord.
Attachment of
the Control Unit
Holder is loose.
Handpiece
Tip oscillates, but
LED does
Handpiece LED turns
not illuminate.
on and off.
(Forza V3 LED)
Loss of the power
Power output
output without
is set 8 at G.
operation.
The AC Adaptor or the
DC Plug is disconnected.
The Tip is not tightened fi rmly.
Worn Tip.
Power has not been correctly
adjusted for the Tip.
The Foot Control is disconnected.
Failure of vibrator in the handpiece.
Failure of internal components of
the Foot Control.
Power has not been
properly adjusted for the Tip.
The Tip is not tightened fi rmly.
Power has not been
properly adjusted for the Tip.
The Tip is not tightened fi rmly.
Failure of vibration in the handpiece
or the Control Unit.
Power has not been
properly adjusted for the Tip.
The Tip is not tightened fi rmly.
Failure of vibration in the handpiece
or the Control Unit.
The Water Adjustment
Knob is closed.
Disconnected Irrigation
supply at low volume range.
(less than 10ml/min.)
The Water Filter is clogged.
O-Ring at the handpiece
cord is worn or damaged.
The holder locking mechanism is
worn out.
The handpiece is not connected into
the Handpiece Cord correctly.
Safety function is activated.
CAUSE
Correctly insert the AC Adaptor
or the DC Plug.
Tighten the Tip until the
Tip Wrench clicks.
Replace the Tip.
Adjust the power level as indicated in the Power
Guide or Tip Case label. Do not exceed MAX.
Connect the Foot Control correctly.
Contact Brasseler USA.
Contact Brasseler USA.
Adjust the power level as indicated in the Power
Guide or Tip case label. Do not exceed.
Tighten the Tip until the
Tip Wrench clicks.
Adjust the power level as indicated in the Power
Guide or Tip case label. Do not exceed.
Tighten the Tip until the
Tip Wrench clicks.
Contact Brasseler USA.
Adjust the power level as indicated in the Power
Guide or Tip case label. Do not exceed MAX.
Tighten the Tip until the
Tip Wrench clicks.
Contact Brasseler USA.
Check the water circuitry and supply to the Control
Unit. Water pressure : 0.1-0.5MPa. (1-5kgf/cm
Turn the Water Volume Knob and adjust to the
appropriate volume.
No problem. Turn the Water Volume Knob and increase
the Irrigation volume.
Replace with new Water Filter. (Refer to 10-3)
Changing Water Filter.
Replace with new O-Ring. (Refer to 10-2 Changing
O-Ring)
Contact Brasseler USA.
Firmly insert the handpiece into the Handpiece Cord.
Powerful output will weaken automatically when used
continously for over 10 minutes at maximum power in
the G mode. Release the foot from the Foot Control.
Decrease the Power less than 5, then increase the
power again. ( Refer to 6-5 )
12
SOLUTION
* Repairs cannot be made by the customer.
)
2