10-3 CHANGING WATER FILTER
Change the Water Filter as necessary.
1) Close the water valve of the Chair Unit.
2) Mount two Spanner Wrenches ( 5x8 ) and turn. (Fig.18)
3) When the Water Filter case is separated, the Water Filter can be removed. (Fig.19)
4) Replace with new (Order Code 5010480U0 ) and reassemble the fi lter in the reverse order.
Control Unit Side
Water Filter Case
11. STERILIZATION
• Autoclave sterilization is recommended.
• Autoclave sterilization required fi rst time you use and after each patient as noted below. Take handpiece out of the packing bag before
sterilization.
• ONLY the Tip (except diamond coated scaler tip), Handpiece, Tip Wrench, Tip Holder and Tip Cover S can be autoclaved. The diamond
coated Scaler Tip is to be sterilized prior to first use, then disposed of as this Tip is single use only.
This handpiece can be cleaned and disinfected with a Thermo-Disinfector.
Autoclave Procedure
1 ) Remove the Tip after use. ( Refer to 5. Mounting and Removing Tip )
2) Clean all the surfaces of the components using a small tooth brush with isopropyl alcohol to remove dirt and debris then dry thoroughly.
3) Insert those into the an autoclave pouch. Seal the pouch. ONLY the Tip (except diamond coated Scaler Tip), Handpiece, Tip Wrench and
Tip Cover S can be autoclaved.
4) Autoclavable up to max. 135˚C.
Type
Minimum Temperature
Full Cycle Time
Minimum Dry Time
5 ) Keep the products in the autoclave pouch to keep it clean until you use it.
Sterilization at 121°C for more than 15 minutes is recommended by ISO17664 and EN ISO17665-1.
The recommended sterilization procedures require the use of FDA-cleared sterilizers, sterilization trays, sterilization wraps, biological
indicators, chemical indicators, and other sterilization accessories labeled for the sterilization cycle recommended.
CAUTION
• Do not sterilize by ultraviolet ray. The handpiece could discolor.
• If autoclaved with other instruments stained with chemical solution, it could strip the plating and make the surface black.
• Do not autoclave the Control Unit, AC Adaptor, Foot Control, Handpiece Cord, O-Rings. Perform alcohol disinfection to the Control Unit,
AC Adaptor, Foot Control, Handpiece Cord after every patient.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
Chair Unit Side
Fig.18
Steam Sterilization Cycle
Gravity Displacement
121° C
132° C
20 Minutes
15 Minutes
10 Minutes
10 Minutes
Control Unit Side
Pre-Vacuum
135° C
10 Minutes
10 Minutes
11
O-Ring
Water Filter
Chair Unit Side
135° C
10 Minutes
15 Minutes
Fig.19