16. SYMBOLS

TUV Rhineland of North America is a Nationally Recognized
Testing Laboratory (NRTL) in the United States and is accredited by
the Standards Council of Canada to certify electro-medical products
with Canadian National Standards.
Consult operation instructions.
Protected against vertically falling water drops.
Type BF applied part.
Class II Equipment.
RX ONLY
U.S. Federal law restricts this device to sale by or on the order of a dentist.
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
The Forza V3/Forza V3 LED is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the
Forza V3/Forza V3 LED should assure that is used in such an environment.
EMISSIONS TEST
RF emissions CISPR11/EN55011
RF emmissions CISPR11/EN55011
Harmonic emissions
EN/IEC61000-3-2
Voltage fl uctuations/fl icker
emissions EN/IEC61000-3-3
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The Forza V3/Forza V3 LED is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the
Forza V3/Forza V3 LED should assure that is used in such an environment.
IMMUNITY TEST
EN/IEC60601 TEST LEVEL
Electrostatic discharge
±6kV contact
(ESD)
±8kV air
EN/IEC61000-4-2
Electrical fast transient/
±2kV for power supply lines
burst EN/IEC61000-4-4
±1kV for input/output
Surge
±1kV line(s) to line(s)
EN/IEC61000-4-5
±2kV line(s) to earth
Voltage dips, short
<5% Ut (>95% dip in Ut)
interruptions and voltage
for 0.5 cycle
variations on power
supply input lines
40% Ut (60% dip in Ut)
EN/IEC61000-4-11
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 secs
Power frequency
3 A/m
(50/60Hz) magnetic fi eld
EN/IEC61000-4-8
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
The Forza V3/Forza V3 LED uses RF energy only for its internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The Forza V3/Forza V3 LED is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Class A
Complies
COMPLIANCE LEVEL
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut(>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
3 A/m
Follow the waste of electric and electronic equipment
(WEEE) Directive (2002/96/EC) to dispose of the product
and accessories.
Marking on the outside of Equipment or Equipment parts that
include RF transmitters or that apply RF electromagnetic energy
for diagnosis or treatment.
Autoclavable up to Max.135°C. * for detail see Sterilization.
Manufacturer.
This product can be cleaned and disinfected with a
Thermo-Disinfector.
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the Forza V3/Forza V3 LED
requires continued operation during power mains interruptions,
it is recommended that the Forza V3/Forza V3 LED be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fi elds should be at levels characteristic
of a typical location in a typical commercial or hospital
environment.