Interferences

Opening the
combination
electrode
20
 several seconds at high concentrations,
 several minutes near the detection limit.
The measured value is stable if the value does not change by more than
0.1 mV within 30 seconds.
Interferences
 Mercury
 Protein

Aging

Please note that every electrode undergoes a natural aging process. The re-
sponse time increases and the slope decreases with the age of the electrode.
The following factors shorten the lifetime considerably:
 Unsuitable storing conditions (see section S
 Special measuring conditions (e.g. organic solutions, frequent measuring
with mercury or protein concentrations)
 High temperatures
 High changes in temperature
The warranty does not cover damage caused by measuring conditions and
mechanical damage.

Maintenance

 Combination electrodes:
Refill any used up bridge electrolyte.
 In the case of increased response times of the electrode perform the fol-
lowing activities:
– Polish the membrane surface using an SF/K polishing strip (see section
R
ECOMMENDED ACCESSORIES
deionized water, and with circular movements polish the membrane
surface for approx. 30 seconds.
– Then recondition the electrode (see section C
,
CALIBRATION
MEASUREMENT
Clean the inside of the combination electrode. To do so, open the
combination electrode as follows:
1
Unscrew the knurled nut from the connection head.
2
Position the connection cable upward in a straight line with the con-
nection head.
TORAGE
). To do so, moisten a polishing strip with
ONDITIONING
).
Ag/S 800
)
,
ba15350d04
12/2021