IKA MF 10 Mode D'emploi page 22

Masquer les pouces Voir aussi pour MF 10:
Table des Matières

Publicité

Mill does not start:
Possible causes:
- Power supply plug is not correctly plugged into the eqip-
ment and/or the socket („Power" LED does not light up if
the rocker switch Pos. A 13 is moved to Position 1).
- Grinding Chamber door not properly locked, („Power"
LED, „Error" LED are lit and beep is heard when the rocker
switch Pos. A 13 is moved to Position 1).
Remedy:
First shift the rocker switch Pos. A 13 to 0. Check voltage supply
and the power supply plug is necessary. Or if „Error" LED is lit
and beeping can be heard close the grinding chamber door and
tighten the turning handle Pos. P 10/ S 9 until stop. The rocker
switch should only then be shifted to Position 1 again.
Engine hums after the rocket switch Pos. A 13 is
shifted but the rotor does not begin to run.
Possible causes
- Ground material chunks is in the grinding chamber before
grinding begins and blocks the rotor.
Remedy:
Switch the mill off at the rockers switch Pos. A 13 and discon-
nect the power supply plug. Open the door and remove the
ground material chunks from the grinding chamber. After the
door is closed and the mill is connected to the mains again, you
can switch the mill on again using the rocker switch Pos. A 13.
Troubleshooting
Engine switches off during operation:
Possible causes:
- Overloading of the engine because of too large feeding
amounts which leads to heating up of the engine. Thus
the motor thermal protective switch is triggered and
switches the mill off. The overheating of the motor is in-
dicated by the red „Error" LED becoming lit and a beep is
heard as well.
- Rotor blocking as a result of feed chunks that are too
large or too tough.
- Rotor blocking due to improper speed adjustment (in general,
the speed is too low for the given application).
Remedy:
Switch the mill off at the rocker switch Pos. A13 and dis-
connect the power supply plug. Open the door and remove
any blocked ground material or ground material that has
been cut up.
If the disturbance occurred because of overloading, let the
mill cool off before you start it again.