Discharge Valve - BRAND seripettor Mode D'emploi

Masquer les pouces Voir aussi pour seripettor:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Cleaning
Cleaning / replacing valves (cont.)
seripettor
pro
®
1.
Filling valve
a) Pull out the filling tube (fig. a).
b) Use the mounting tool to unscrew the filling valve (fig. b).
c) If the sealing ring is contaminated or damaged, carefully
remove it with a pair of curved forceps (fig. c).
d) If necessary, clean any contaminated individual parts
(e.g., in an ultrasonic bath).
e) Insert cleaned or new sealing ring.
f) Screw in the cleaned or replacement filling valve, first by
hand, and then tighten it with the mounting tool (fig. f).
2.

Discharge valve

The discharge valve is integrated in the discharge tube.
a) Remove the discharge tube and clean it in the ultrasonic bath
if necessary.
b) Mount cleaned or new discharge tube (s. page 33, fig. 1').
Note:
If the instrument does not fill up, and if some elastic resistance is
evident when the piston is pulled upward, then it is possible that the
ball valve is merely stuck.
In this case, loosen the ball valve using light pressure, for example,
with a 200 µl plastic pipette tip (see fig.).
42
a
b
c
f

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Seripettor proSeripettoSeripetto pro

Table des Matières