11 - TROUBLE-SHOOTING CHART
PROBLEMS POSSIBLE
CAUSES
• No air or
• Flow generator
insufficient
off
air comes
• Flow generator
out of the
regulated
mask
wrongly
• Patient tube not
connected
• The mask is
• Mask
noisy
assembled
incorrectly
• The mask
• Mask not
does not
positioned
fit the face
correctly on the
snugly
patient's face
• Headgear
straps too tight
ENGLISH
SOLUTIONS
• Switch on the flow
generator.
• Regulate the flow
generator
• Connect the patient tube
to the flow generator and
mask.
• Take the mask apart
and reassemble it, as
explained in section HOW
TO ASSEMBLE THE
MASK.
• Reposition the mask, as
explained in section HOW
TO PUT THE MASK ON
THE CHILD.
• Unfasten the lower straps
and refasten more loosely.
12 - DISPOSAL INSTRUCTIONS
The components of the mask are made of the following
materials and must be disposed of in accordance with the
applicable laws.
COMPONENT
Main body (a)
Tube assembly (b)
Headgear (c)
13 - TECHNICAL DATA
Normative references CE marking in accordance with
Dead space volume
inside the mask
Operating pressure
Temperature range
Max. overall
dimensions
14 - WARRANTY
Air Liquide Medical Systems S.r.l. warrants the product free
from material and manufacturing defects for a period of
120 days from the date of purchase, provided the operating
conditions and instructions are complied with.
If the product is found to be faulty during normal use, it will be
repaired or replaced at the manufacturer's discretion.
The manufacturer accepts liability for the safety, reliability and
15
MATERIAL
Silicone + Polycarbonate
Polycarbonate + Metallocene
+ Acetalic + Silicone
Nylon + polyurethane
European Directive 93/42/EEC, class
IIa, as amended, EN ISO 14971, ISO
5356-1, ISO 10993-1 and EN ISO
17510-2, EN ISO 17664
38 ml
4-24 cmH
O
2
5-40°C
86 mm x 70 mm x 222 mm