9 Troubleshooting

9.1 Identifying and repairing faults

The following table provides information on malfunctions and their possible causes. If you cannot correct the malfunction
using the remedies named, please contact your dealer directly.
Fault
No motor noise audible/lift
does not move.
The lift can be raised but not
lowered.
The lift stops while being
raised.
The bath lift is not adhering
securely to the bathtub.
Battery is not charging and
the red indicator light flashes
during charging
1574470-E
Possible cause
Jack socket of the hand control
is not engaged correctly.
Battery is flat.
Contacts are dirty.
Cable is crushed/damaged.
Low-voltage protection is
activated.
Excessive load.
Suction cups are old or
damaged.
Faulty electronics.
Faulty battery.
Remedy
Engage the jack socket, → 4.2 Setting up Bath Lift, page 11.
Recharge the battery completely, → 4.3 Charging Battery,
page 15.
Clean the plug on the backrest, → contact your dealer, if
necessary.
Replace the hand control, → contact your dealer, if necessary.
Recharge the battery completely, → 4.3 Charging Battery,
page 15.
Observe maximum load, → 10.1 Dimensions and weight, page
27.
Replace the suction cups, → contact your dealer, if necessary.
Have your authorized dealer check the hand control.
Replace the hand control.
Troubleshooting
25