8 Troubleshooting

8.1 Identifying and repairing faults

The following table provides information on malfunctions and their possible causes. If you cannot correct the malfunction using the
remedies named, contact your dealer directly.
Fault
No motor noise audible/lift does
not move.
The lift can be raised but not
lowered.
The lift stops while being raised.
The bath lift is not adhering
securely to the bathtub.
Battery is not charging and the
red indicator light flashes during
charging.
LED lights up while lifting the bath
lift.
1584939-A
Possible cause
Jack socket of the hand control is
not locked correctly.
Battery is flat.
Contacts are dirty.
Cable is crushed/damaged.
Low-voltage protection is
activated.
Excessive load.
Suction cups are old or damaged.
Faulty fuse.
Faulty battery.
Faulty charger.
Battery nearly empty.
Remedy
Lock the jack socket. See 3.3 Setting up the bath lift, page 9 .
Recharge the battery completely. See 3.4 Charging the battery, page 11.
Clean the plug on the backrest. Contact your dealer, if necessary.
Replace the hand control. Contact your dealer, if necessary.
Recharge the battery completely. See 3.4 Charging the battery, page 11.
Observe maximum load. See 9.1 Dimensions and weight, page 22.
Replace the suction cups. Contact your dealer, if necessary.
Have your authorized dealer check the hand control.
Replace the hand control.
Have your authorized dealer check the charger.
Recharge the battery completely. See 3.4 Charging the battery, page 11.