Administrative Mode
The factory default settings of the Invia Liberty pump are -125 mmHg and constant mode.
CAUTION
The pressure level should always be set according to prescribing healthcare professional instruction.
When an Invia FitPad is used, the set pressure is controlled at the wound site.
Device setup
New patient therapy
1. Switch Invia Liberty pump ON in Administrative mode,
press and hold [
], and then press [
Self-test starts
If the self-test is not successful, follow instructions shown on
the display or see the Alarm chapter.
2.
Acknowledge disclaimer. Press
"New patient? Yes / No"
"Yes"
means that Invia Liberty pump will issue a new therapy
ID number (= new patient).
This number can be noted in the patient's file.
The therapy ID number is displayed in the information field.
If yes
Press
"Yes"
[
] to confirm.
Press
"OK"
[
] to enter the main display.
"No"
means that the therapy ID number and settings remains
unchanged (= same patient).
If no
Press
"No"
[
] to confirm and enter the main display.
].
"OK"
[
] to confirm.
19