Cleaning and disinfection

CAUTION
Invia Liberty pump with associated products (docking station, rail holder and charger) should be
cleaned / disinfected after every use. Before cleaning the device, unplug the pump from the wall
outlet.
Pump
Docking station
Rail holder
Wipe off with
Charger
a damp cloth.
Carrying case*
* If needed, the carrying case can be washed in washing machine
CAUTION
Invia Liberty pump, docking station, rail holder, charger and carrying case cannot be sterilized.
Immersion disinfection, thermal disinfection and ultrasound cleaning are not permitted.
Disinfection (Pump housing, docking station and charger)
CAUTION
Invia Liberty pump can be disinfected with "alcohol".
Disinfection
Invia Liberty can be disinfected with the disinfecting agent group "alcohol".
Do not use other cleaning agents (e.g. Terralin) as they can damage the plastic housing.
Immersion disinfection, thermal disinfection and ultrasound cleaning are not permitted.
Sterilization
Invia Liberty and Invia Liberty accessories cannot be sterilized.
CAUTION
Do not use other cleaning agents (e.g. Terralin) as they can damage the plastic material.
Cleaning procedure for Invia Liberty pump, docking station and charger
1. Wear suitable protection (clothing, gloves, face mask and goggles) according to local guidelines.
2. Apply the disinfectant agent in accordance with instructions from the manufacturer
Pay attention particularly to edges, narrow corners and bottom side.
3. Leave the disinfectant on. Follow the recommended residence time for the disinfectant
as instructed by the manufacturer
4. Thoroughly clean the surface, all edges, housing niches, corners, bed holder, brackets,
port covers and bottom side.
38
Cleaning
Disinfection
P
P
Wipe off with a dis-
infecting agent
.
1)
Sterilization
O
Sterilization / cleaning in a dishwasher is
NOT allowed.
.
1)
Dishwasher