ANALOGUE PROBE CALIBRATION (S75...)
ENTER
2.0 CALIBRATION
View settings
Edit settings
Reset Defaults
Span
UP
DOWN
2.0 CALIBRATION
View settings
Edit settings
Reset Defaults
Span
UP
DOWN
2.0 CALIBRATION
View settings
Edit settings
Reset Defaults
Span
2.0 CALIBRATION
View settings
Edit settings
Reset defaults
Span
This programming step allows for the calibration of the instrument/measurement probe.
A "standard" calibration curve has been prepared for each of the probes that can be connected to the
ACP 4063. These curves are comprised of 8-point tables drawn up under laboratory conditions
using Silica or Formazin solutions of known concentration, where values in milliVolts (signal
generated by the measurement probe) are associated with concentration values expressed in the
various units of measure. Once the probe typology has been inserted (sec. 4.3.3. C 4), the system
will use the corresponding "standard" calibration curve by default.
For this reason, the measurement system will not require calibration up start up.
0000137221
2323 mV
0mg/l
2145 mV
10000mg/l
2056 mV
20000mg/l
1957 mV
30000mg/l
1843 mV
40000mg/l
1755 mV
50000mg/l
1666 mV
60000mg/l
1577 mV
70000mg/l
2323 mV
0mg/l
2145 mV
10000mg/l
2056 mV
20000mg/l
1957 mV
30000mg/l
1843 mV
40000mg/l
1755 mV
50000mg/l
1666 mV
60000mg/l
1577 mV
70000mg/l
2.3 CALIBRATION
Edit stinga
Default values
Are you sure?
2.4 CALIBRATION
Span
Rev. 3.1