Accessing the Clinician Menu

Impedances

You can use the remote control to check impedances.
From the Clinician Menu:
Impedances. An impedance measurement is taken and the Impedances screen is
1. Select
displayed.
When an impedance measurement is taken, impedances are assessed between a contact and the
case (monopolar), and between pairs of contacts (bipolar). A green dot indicates that impedance is
within the acceptable range. A yellow dot with a question mark indicates that impedance is outside of
the acceptable range.
Monopolar impedance values can be displayed by pressing the Down
Arrow button from the
Impedances screen.
2. Select Measure
to check impedances again.
3. Select
Back to go back to the Clinician Menu.
Vercise™ DBS Remote Control 2 Directions for Use