Press
to clear the message. The message will reappear approximately weekly until the end of the
Stimulator's programmed period is reached. When the end of its programmed period is reached, the
remote control will display the following on-screen message.

ERI (Elective Replacement Indicator) Screen

The implanted non-rechargeable Stimulator is nearing end of service. Changes made to the stimulation
will not be saved, and stimulation will not be available soon. Patients should be advised to contact their
physician to report this message screen. Press the select key to clear this information screen. This
screen will appear whenever the remote control is activated.
Vercise™ DBS Remote Control 2 Directions for Use
90985317-02 REV A 22 of 557