1.2

Technical specifications

(Given in fully assembled condition)
Make
Address
Type
Model
Description
Maximum occupant mass
Length
Width
Height adjustment of the slatted
frame
Total mass (excl. side rails,
lifting pole and triangle)
Adjustment of the head end
Max. thigh angle
Max. angle for lower leg
Max. safe working load *
Lifting pole
Mattress
Under clearance for patient
hoist
Steering casters (4 pieces)
Motor undercarriage
Head end motor
Feet end motor
Handset
Transformer
Nominal voltage, transformer
Period switched on
Noise level
Class / Kind of protection
Storage and use temperature
Storage and use humidity
We reserve the right to introduce technical changes. Measurement tolerance ± 15 mm / 1,5 kg / °
The bed complies to the requirements set up in:
EN 1970:
Adjustable beds for disabled persons: Requirements and test methods
EN 1970+ A1:
Adjustable beds for disabled persons: Requirements and test methods
EN 60601-2-38/A1:
Medical electrical equipment - Part 2-38: Particular requirements for the
safety of electrically operated hospital beds
Vermeiren
Vermeirenplein 1/15, B-2920 Kalmthout
Bed
Pratic HC
Min.
Continuously from 400 to 800 mm (without mattress)
Min. RG 35, thickness 120 mm, 2000 x 900 mm, max. 20 kg
Ø 50 mm, with central brake
U input: 230 V, ± 10%, 50 Hz, current: max. 1.5 A,
output voltage: 24V
10%, Maximum two minutes then 18 minutes for non-use
Protection class II, application part Type B / IPX4
+ 5 °C
30%
* Max. safe working load = mass user weight + mass mattress + mass other accessories
Table 1: Technical specifications
Page 4 
Max.
Nominal load 135 kg
2070 mm
1050 mm
98 kg
70°
30°
-16°
170 kg
Nominal load 80 kg
150 mm
Linak LA27
Linak LA27
Linak LA27
Linak HB7x, IPX4
Linak CB6
, max. 70 VA
~ 48 dB(A)
+ 40 °C
Pratic HC
2014-06
(spray protection)
70%

Publicité

Table des Matières
loading