INSTALLATION / USE
The X-Cid
®
must be installed in a well ventilated room. The air exchange rate must be at least 100m
per device. The air exchange may be passive or may be assisted by static extractors, air conditioners, or
mechanical extractors. The technician will provide the end user with advice in this regard.
The X-Cid
is designed to be installed on a horizontal surface, but may be wall mounted using the
®
following method:
• Make sure the wall can support the weight of the device (approx. 8 kg).
• Place two anchors (rawl plugs) for Ø 5 to 5.5 mm screws in the wall 200 mm apart. The screws must be
screwed into the anchors so that their heads (Ø 10 maxi) protrude 7 mm from the wall.
Empty the reservoirs and remove the 500 I CO2 spray can before moving the device.
If the X-Cid
®
is not used for a year or more, it should be inspected by a certified installer.
EQUIPMENT
Equip each station with the following parts, depending on the instruments to be treated:
• ISO 3694 coupling (contra-angles and handpieces).
• ISO 9168 MID coupling (turbines) with suitable quick-connector.
Compressed air connection (1)
The device is equipped with a quick-connector for Ø 4x6 flexible tubing.
Check that the tubing has a clean cut to ensure an airtight connection.
Electrical connection (2)
Plug in the X-Cid
®
using the power cord provided. Comply with the safety data sheet.
The wall socket must be easily accessible and must be grounded.
Fluid evacuation (3)
The tubing for evacuating waste fluids (Ø 25 ext.) is directly connected to the waste outlet. All components
are biodegradable.
Make sure that the connection is watertight and that the tubing has a sufficient downward slope.
2
16
3
/h
3
1

DIRECTIONS FOR USE

1) Make sure that the fluids exceed the minimum required levels.
2) Turn on the device by pressing the 0/1 button. The green "power
on" LED will light up.
3) Open the hinged door of the treatment tank and install the
instruments to be treated:
• Insert contra-angles and handpieces completely. You will hear
a click. To remove them, press the button and pull.
• Place turbines on the quick-connect coupling.
4) Close the door.
5) Press buttons 1, 2, and/or 3 to activate the corresponding
treatment stations. The yellow LEDs of the selected stations
will blink.
• Caution ! Instruments should undergo another complete treatment cycle in the event the cycle is
voluntarily or involuntarily interrupted in order to ensure an effective treatment.
• Only select those stations with an attached instrument. Otherwise, cleaning product will be wasted.
• Make sure the stations to be used are correctly selected on the keypad.
6. Press the "start cycle" button.
• During the cycle, the yellow "cycle" and "stations in use" LEDs will light up.
• At the end of a cycle, the "cycle" LED will turn off and the "stations in use" LEDs will blink.
• If the user voluntarily stops the cycle, the "stations in use" LEDs will turn off (which indicates a reset). The
user must reselect the stations before starting a new cycle.
Non-lubricating mode
You may also operate the X-Cid
in non-lubricating mode by pressing the "1" button at the same time as the
®
"power on" button.
If the non-lubricating mode is selected, the red LED will light up for 5 seconds when the power is turned on.
X-Cid
®
keypad in
non-lubricating mode.
17