MAINTENANCE AND TIPS
The X-Cid
has been designed to clean and lubricate handpieces, contra-angles, and turbines after
®
each patient.
The following tips will help you make the best use of the device and your instruments:
If your instruments are not new, we recommend a complete internal cleaning to remove accumulated
debris prior to a treatment with the X-Cid
.
®
Treat your instruments immediately after use so that saliva and blood do not have time to coagulate or dry.
Once treated, instruments must be packaged for sterilization.
During long dental procedures (> 15 min) with fast rotating instruments (contra-angles with red rings or
turbines), we recommend that you manually lubricate your instruments during the procedure because the
air flow tends to eliminate the lubricant.
Daily maintenance
Before the first daily use of the X-Cid
®
, verify the fluid levels. In the event you forget this step, the cycle will
stop automatically and an alarm will be triggered when fluid levels fall too low.
Clean the outside of the device with a soft towel dipped in disinfectant. Never use abrasives.
Monthly maintenance
Select all three stations, operate the device without instrument, and verify that there is a good flow of fluid
at each station.
Annual maintenance
The volumetric flow of the device must be tested one a year by a manufacturer-certified technician.
Fuses
The fuses are located near the ON/OFF switch. They must be replaced by fuses of the same type (see
Characteristics).
CHARACTERISTICS
Definition:
The X-Cid
is an automated pre-sterilization cleaning device.
®
• Ambient temperature for optimal operation: 10°C to 40°C (50°F to 104°F)
• Ambient humidity for optimal operation: 10% to 80% RH
• Atmospheric pressure for optimal operation: 500 hPa (millibars) to 1,060 hPa (millibars).
Dimensions:
H=47cm /L=40cm /W=23cm
Weight:
8 kg
Electrical:
Ref.: 51100003 = 230V +/- 10% 50-60 Hz EU
Ref.: 51100004 = 230V +/- 10% 50-60 Hz CH
Ref.: 51100005 = 230V +/- 10% 50-60 Hz UK
Ref.: 51100007 = 110V +/- 10% 50-60 Hz USA
Ref.: 51100009 = 100V +/- 10% 50-60 Hz Japan
Electric class: I
Power:
38 Watts
Protection :
• Two 0.2 Amp fast blow fuses (5x20 mm) for the 230 V X-Cid
• Two 0.5 Amp fast blow fuses (5x20 mm) for the 100 and 110 V X-Cid
18
®
®
Compressed air feed:
• Feed pressure: max.=8 bars / min.=5 bars
• Average flow: 2 l/minute
• Air quality: 1 micron filter; residual water/oil=0.5 ppm at 25°C
• Connection: quick-connect for 4x6 tubing
Evacuation:
Tube (Ø 25 mm ext.) connected directly to the waste fluid outlet without siphon.
Treatment fluids:
• Pre-disinfection and cleaning: X-Cidol
®
2 detergent; Ref.: 50704005
• Rinsing: Demineralised water
• Lubricating: MICRO-MEGA
500 I CO2 oil; Ref.: 50700029
®
INSTRUMENTS TREATED
The X-Cidol
2 detergent solution penetrates the components and structures of all commercially available
®
handpieces, contra-angles and turbines, from the bottom to the orifice at the head.
In case of any doubts, contact your distributor.
It is essential to carefully examine your old instruments before treating them with the X-Cid
The manufacturer reserves the right to modify the construction and characteristics of the device without
prior notice.
WARRANTY
For any manipulations other than those described above, please contact the service department of your
distributor.
The X-Cid
is guaranteed for 24 months following the installation date against any manufacturing defects,
®
with the exception of damage caused by improper handling or maintenance, the use of unsuitable products
or products that are not recommended by the manufacturer (detergent/disinfectant, demineralised water
and oil), or non-compliance with the operating instructions for the device.
.
®