MAINTENANCE AND TIPS
The X-Cid
has been designed to clean and lubricate handpieces, contra-angles, and turbines after
®
each patient.
The following tips will help you make the best use of the device and your instruments:
If your instruments are not new, we recommend a complete internal cleaning to remove accumulated
■
debris prior to a treatment with the X-Cid
.
®
Treat your instruments immediately after use so that saliva and blood do not have time to coagulate or dry.
■
Once treated, instruments must be packaged for sterilization.
■
During long dental procedures (> 15 min) with fast rotating instruments (contra-angles with red rings or
■
turbines), we recommend that you manually lubricate your instruments during the procedure because the
air flow tends to eliminate the lubricant.
Daily maintenance
Before the first daily use of the X-Cid
®
, verify the fluid levels. In the event you forget this step, the cycle will
■
stop automatically and an alarm will be triggered when fluid levels fall too low.
Clean the outside of the device with a soft towel dipped in disinfectant. Never use abrasives.
■
Monthly maintenance
Select all three stations, operate the device without instrument, and verify that there is a good flow of fluid
■
at each station.
Annual maintenance
The volumetric flow of the device must be tested one a year by a manufacturer-certified technician.
■
Fuses
The fuses are located near the ON/OFF switch. They must be replaced by fuses of the same type (see
■
Characteristics).
CHARACTERISTICS
Definition:
The X-Cid
is an automated pre-sterilization cleaning device.
®
• Ambient temperature for optimal operation: 10°C to 40°C (50°F to 104°F)
• Ambient humidity for optimal operation: 10% to 80% RH
• Atmospheric pressure for optimal operation: 500 hPa (millibars) to 1,060 hPa (millibars).
Dimensions:
H=47cm /L=40cm /W=23cm
Weight:
8 kg
Electrical:
Ref.: 51100003 = 230V +/- 10% 50-60 Hz EU
Ref.: 51100004 = 230V +/- 10% 50-60 Hz CH
Ref.: 51100005 = 230V +/- 10% 50-60 Hz UK
Ref.: 51100007 = 110V +/- 10% 50-60 Hz USA
Ref.: 51100009 = 100V +/- 10% 50-60 Hz Japan
Electric class: I
Power:
38 Watts
Protection :
• Two 0.2 Amp fast blow fuses (5x20 mm) for the 230 V X-Cid
• Two 0.5 Amp fast blow fuses (5x20 mm) for the 100 and 110 V X-Cid
18
®
®
Compressed air feed:
• Feed pressure: max.=8 bars / min.=5 bars
• Average flow: 2 l/minute
• Air quality: 1 micron filter; residual water/oil=0.5 ppm at 25°C
• Connection: quick-connect for 4x6 tubing
Evacuation:
Tube (Ø 25 mm ext.) connected directly to the waste fluid outlet without siphon.
Treatment fluids:
• Pre-disinfection and cleaning: X-Cidol
®
2 detergent; Ref.: 50704005
• Rinsing: Demineralised water
• Lubricating: MICRO-MEGA
500 I CO2 oil; Ref.: 50700029
®
INSTRUMENTS TREATED
The X-Cidol
2 detergent solution penetrates the components and structures of all commercially available
■
®
handpieces, contra-angles and turbines, from the bottom to the orifice at the head.
In case of any doubts, contact your distributor.
■
It is essential to carefully examine your old instruments before treating them with the X-Cid
■
The manufacturer reserves the right to modify the construction and characteristics of the device without
■
prior notice.
WARRANTY
For any manipulations other than those described above, please contact the service department of your
■
distributor.
The X-Cid
is guaranteed for 24 months following the installation date against any manufacturing defects,
■
®
with the exception of damage caused by improper handling or maintenance, the use of unsuitable products
or products that are not recommended by the manufacturer (detergent/disinfectant, demineralised water
and oil), or non-compliance with the operating instructions for the device.
.
®