INSTRUCTION MANUAL
This device has been designed and manufactured in accordance with EU Directive 93/42/CE.
X-Cid
®
is an automated device for cleaning handpieces, contra-angles, and turbines.

SAFETY INSTRUCTIONS

X-Cid
must only be used by qualified personnel to pre-disinfect handpieces, contra-angles, and turbines.
®
The instrument must be in good working order. It may not be possible to properly clean a damaged
instrument.
Always wear clean, disposable medical gloves when handling handpieces, contra-angles, turbines and X-Cidol
Only use manufacturer recommended products with the X-Cid
• X-Cidol
2 detergent; Ref.: 50704005
®
• Demineralised water
• 500I CO2 MICRO-MEGA
oil; Ref.: 50700029
®
• The device must be installed by a manufacturer-certified technician following the instructions in the
installation manual.
Do not store the device near a heat source.
Do not expose the X-Cid
to direct sunlight.
®
Do not smoke in close proximity to the device.
The air exchange rate in the room where the device is installed must be at least 100m
In the event the device is not used for four (4) weeks, complete an empty cycle before installing instruments
in the treatment tank.
If detergent comes in contact with the eyes, rinse immediately and thoroughly with water and consult a
physician. Consult the X-Cidol
®
2 safety data sheet available from your distributor or MICRO-MEGA
Do not disassemble or damage the device.
No not place your hands on the hinges when opening the window.
Make sure the device is solidly seated before use.
The wall socket must be grounded.
To unplug the device, remove the power cable.
Unplug the device and contact your distributor:
• If the device seems to be operating abnormally (time cycles not respected, abnormal liquid consumption,
abnormal noise, etc.)
• If the liquid is no longer evacuated from the reservoir.
• If there is a leak at the back of the device.
• If there is visible damage to the device.
Contact your distributor or MICRO-MEGA
in order to dispose of the device at the end of its service life in
®
accordance with EU Directive 2002/96/CE .
GENERAL
Introduction
The X-Cid
®
automatically pre-disinfects, cleans and lubricates handpieces, contra-angles, and turbines
with no disassembly required.
Treated instruments are clean and in perfect working order, ready to be packaged and sterilised.
Since the X-Cid
complies with Administrative Circular N° DGS/5C/DHOS/E2/2001/138 of 14 March 2001,
®
patients and health personnel can be assured that there is no risk of cross-infection. In addition, the X-Cid
increases the service life of instruments, reduces maintenance costs, and is compatible with all brands of
handpieces, contra-angles, and turbines.
12
0459
.
®
®
:
/h.
3
®
.
®
PROCEDURE
Pre-disinfection (15 min)
All surfaces in contact with pathogens are treated with X-Cidol
disinfectant solution that complies with current standards. The internal parts of instruments are
rotated at low speed to eliminate interfering substances as well as most microorganisms.
Rinsing (2 min 30 s)
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
the previous phase.
Cleaning (5 min 30 s)
The instruments are again rotated in X-Cidol
phase and eliminates all remaining microorganisms, leaving the instruments totally clean.
Rinsing (2 min 30 s)
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
the previous phase.
Drying (3 min)
Once clean and rinsed, the instruments are dried using compressed air.
Lubricating (2 min 45 s)
The rotating instruments are then lubricated using special oil applied using a mounted spray can.
The excess lubricant is removed using compressed air.
DESCRIPTION
The device is mounted on a stainless steel frame with a self-extinguishing synthetic hull and includes:
A product
reservoir
A keyboard with the
control keypads and
indicators
A stainless steel
treatment tank with a
window
®
2, an aldehyde-free detergent/
2 used in
®
2. This phase complements the pre-disinfection
®
®
2 used in
Two (2) standard ISO 3964 couplings
for handpieces and contra-angles
and one (1) ISO 9168 (MID) coupling
for a turbine
A dock for the
oil spray can
Electrical, compressed air, and
fluid evacuation outlets on the side
of the steel frame
13