1) Treatment Tank:
The treatment tank, which has a window, has
■
three stations that can accommodate instrument
couplings in any order. Each coupling is held in
place by two screws (1). Two types of couplings
can be used:
• Type ISO 3964 for handpieces and
contra-angles (2).
• Type ISO 9168 (MID) for turbines (3); these
couplings can be equiped with quick-
connectors for each turbine.
The order of the coupling can be changed to
■
accommodate the types of instruments being
used.
Each station is equipped with four spray
■
nozzles to treat the external surfaces of the
instruments (4).
1
2
2) Product reservoir:
The reservoir is located in the top of the device and is composed
■
of two compartments with independent hinged lids. It has been
designed for use with the following products:
• X-Cidol
®
2 detergent / disinfect for the
pre-disinfection and cleaning phases.
• Demineralised water for the rinsing phases.
To fill, the covers of the two compartments are
■
opened and the appropriate quantity of product
is added up to the fill line using a funnel, if
necessary.
Do not exceed the fill line to avoid overflows.
■
14
station 1
station 2
station 3
4
3
3) Oil spray can:
The MM 500 I CO2 oil spray can*, with its cap removed, is placed in its dock on the right hand side of the
■
device. A set of sensors monitors the correct positioning and pressure of the spray can.
4) Keypad:
The control keypad includes the following components:
■
• A cycle start button
• Three buttons to select the desired stations
• Green, yellow and red LED indicators:
- Power on (green)
- Stations selected and cycle underway (yellow)
- Malfunction (red).
Remove cap from
spray can.
Place oil spray can
in its dock.
* Use only MICRO-MEGA
®
500I CO2 oil spray cans
(C02 propellant)
LED cycle
Cycle start
underway
button
Station 1
button
LED
power on
Station 2
button
LED alarm
Station 3
button
X-Cid
Keypad
®