Commissioning, Operation - WIKA TR10-0 Mode D'emploi

Type de protection contre l'ignition fp (ex d) et xp (classe/division)
Masquer les pouces Voir aussi pour TR10-0:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5. Commissioning, operation

5. Commissioning, operation
5.1 Removal and installation of the measuring insert
EN
When servicing or recalibrating is necessary, make sure that both surfaces of the flameproof
joint (flame path fitting and measuring insert) are not damaged when installing the
measuring insert. Damaged or scratched parts of the flameproof joint must be replaced.
5.2 Electrical connection
5.2.1 Connection to terminal block
Model TR10-0
1 x Pt100, 2-wire
white
2 x Pt100, 2-wire
black
white
Assignment and colour coding for Pt1000 as well as for Pt100
Pt1000 only available as single elements
10
1 x Pt100, 3-wire
red
white
red
white
2 x Pt100, 3-wire
black
yellow
red
white
red
white
black
yellow
WIKA operating instructions models TR10-0, TC10-0 (FM, CSA)
1 x Pt100, 4-wire
white
red
red
white
red
red
white
2 x Pt100, 4-wire
black
yellow
white
red
red
red
white
black
black
yellow
red
red
red
red
white
white
yellow
red
red
red
white
white
black
black
yellow
yellow

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Tc10-0

Table des Matières