4.1.2
Measurement
Attach the selected rotor to the drive shaft.
Display:
Action:
- Short press the main switch to start the measurement.
"Rx wait'' is displayed until a defined measuring condition is reached (about
3 seconds). The values recorded thereafter will be used to establish an
average value for the viscosity. Finally, the result will be displayed showing
the selected rotor (Rx) and "n" the viscosity value in mPas.
By pressing the main switch the measurement will be started in the
selected range, the motor turns continuously.
Display:
Action:
- Short press the main switch to stop the measurement.
The last measuring value calculated for the viscosity will be "frozen".
By pressing the main switch again the measurement will be restarted in the
range last used.
Error Messages:
Display:
Value lower than minimum measuring range.
Recommendation: Stop the measurement and use a bigger rotor. See section 6.
for rotor dimensions.
Display:
Value higher than maximum measuring range.
Recommendation: Stop and use the optional smaller rotor.
14
Start
Rx
n mPas
Rx
n ! mPas
Rx
>n mPas
SKF TMVM 1