• Rotor selection
Display:
Rx: x is the latest selected rotor before the last switch off, i.e. 1,2 or 3.
The rotor selection is preset with R3.
The last used rotor selected is saved when the viscometer is switched off.
The rotor selection mode allows change of the measuring rotor for different
viscosity measurement range, see section 2.3.
TMVM 1 is supplied with rotor 3 only. Others rotors are optional.
Action:
- Normal operation: when a new rotor selection is not required, wait 3
seconds until the Measurement mode is automatically reached.
- New rotor selection: Short press the main switch within 3 seconds.
The display will then show...
Display:
Rotor selection procedure:
Rx goes from R1, R2 to R3
Action:
- Press the main switch when the required rotor number is displayed.
Display:
SKF TMVM 1
Sensor Rx
Select sensor Rx
Rx selected
13