- Green: the average (residual) charge measured by the sensor bar.
If the screen is clicked somewhere, the screen will "freeze", and display On hold. Clicking the
screen again will display the current value.

7.2.3 Action log tab

In the Action log tab, the changing status of the bar is logged step-by-step. This is indicated
including the date and time the status change has occurred. Main messages:
- HV OK: the bar is turned on and is functioning properly
- HV NOT OK: the bar is turned off or the HV doesn't function properly
- CLEAN BAR: the efficiency has dropped below the clean bar warning level.

7.2.4 Data log tab

In the Data log tab, all available measurement data of the bar is logged at fixed times.
If the parameter Datalogging is activated, these values will also be saved in the log file of the
Manager. The interval between logging is determined by the setting in the Manager IQ Easy.
See user manual of the Manager in Chapter 6.6.3
Main measurement values:
- U- : the output voltage of the negative high-voltage.
- U+ : the output voltage of the positive high-voltage.
- I- : the output current of the negative high-voltage.
- I+ : the output current of the positive high-voltage.
- Eff: the efficiency of the bar

7.3 Functional check using the LEDs (connected to 24 V power supply (not IQ)

The colours of the status LEDs on the anti-static bar indicate the status of the anti-static bar.
Table 2, overview LED indication anti-static bar
LED indication
Green flashing 1 Hz
Green on
Orange
Red on
Red flashing 1 Hz
* The high voltage and signal HV OK can be on or off, depending whether the bar is active (Remote ON) or on
standby (Remote OFF).

7.4 Causes for overloading

Causes for overloading of the high voltage may be:
- Fouling of the anti-static bar.
- Direct contact between conductive material and emitters.
The anti-static bar is mounted too close to conductive (machine) parts.
Causes for overloading of the HV OK signal may be:
Short circuit of the HV OK signal.
- Overload on the HV OK signal.
PerformaxIQEasyPerformaxIQEasySpeed_UM_9752154110_NL_D_GB_FR_V1_5
Status
Standby
In operation
Clean bar
High-Voltage overload
HV OK signal overload
High voltage
Signal HV OK
Off
On
On/Off*
regulated
On
Off
On
On/Off*
Off
Off
GB
71