Final inspection
Inspect Helena for signs of damage. Check to ensure that Helena is standing securely on the floor, so that it does not lean,
and that the support legs are adjusted to the same height.
Check to ensure that all locking knobs and bolts are securely tightened.
The lift unit must be inspected according to the procedure for periodic inspection for the respective lift unit.
Keep the manual where it is easily accessible for users of the product.
Using the product
Important information
• Helena must be assembled according to the assembly instructions provided with the stand.
• Helena may only be used indoors and on a level floor.
• Lifting accessories must be properly trial fitted and tested in relation to the patient's needs and functional ability.
• Do not leave the patient unattended during a transfer situation.
• Under no circumstances may max. load be exceeded. See section on safe working load.
• To ensure optimal function, Helena must be inspected regularly. See section on maintenance.
• Warranty applies only if repairs or alterations are done by an authorized technician.
• Ensure that secondary rail overhang does not exceed 50 mm! Risk for instability!
• Maximum height must not be exceeded! Risk for instability!
• Make sure the longer bolt (M10x25) is inserted into the safety hole when mounting the primary and secondary rails.
Safe working load
Different products on the same lift system (lift unit, slingbar, sling, scales and other lifting accessories) may have different
allowable safe working loads. The lowest allowable safe working load always determines the safe working load of the
assembled system. Always check the safe working loads for the lift and accessories before use. Contact your dealer if
you have any questions.
8