Max 50mm!
-
NOTE! Ensure that secondary rail overhang does not exceed 50mm!
Risk for instability! (Image 7b)
-
NOTE! Make sure the long bolt (10) is inserted into the secondary
rail's safety hole (image 7c).
Finally, mount the following on the secondary rail in the following order:
- End stop
- Lift unit with carriage
- End stop
- Screw covers on all visible bolt heads
Adjust Helena to the desired height by first loosening the locking knobs on the support leg pairs
(one pair at a time) and then carefully rocking the lift stand.
Mount the locking knobs in appropriate holes.
-
when the desired height has been set. Risk for instability! (Image 8)
8
I F U
7b
NOTE! Maximum height must not be exceeded! The red warning label must not be visible
Mount secondary rail on traverse rail carrier assembly
- Insert the rail over the square nuts on the
traverse rail carrier. Tighten with a 17mm
open-end wrench (image 7).
7
M10x25
M10x25
M10x16
M10x16
7c
7