– leave configuration menu.
*
1
filter value='disp' means that the analogue output is derived directly
from the display value. The filter is adjusted for the display but it also
affects the analogue output. The accuracy of the analogue output
depends on the calibration of the display reading. Filter value > 0
means that the analogue output is independent of the display filter.
The results continue to be displayed, with the maximum possible
accuracy, independent of the calibration of the display. At small
values, the analogue output reacts faster to changes of the input
signal. However, the analogue output signal becomes unsteady. At
larger values, the analogue output signal is "smoothed" and thus
steadier – the changes of signal may be delayed.
Important
If analog output 0 – 10 V has been selected, a 500 : resistor must be
connected between pin 3 and 4.
4.6

Configure output function

Hint
The configuration of the measurement input can affect the configuration
of the output functions. Therefore make sure that the output, switching
and alarm are configured a a fter the measurement input.
The analog output can be freely adjusted within the limits of the display
range.
1.
Switch the instrument on.
2.
Wait until the segment test is terminated –
3.
(key 1) + key 5 (rear side) > two seconds –
18
set D/A-converter-filter (filter of the analog output): diSP 1 to
1
.
.
.
select desired output state in case of an error:
– inactive in case of error, output signal 0 mA / 0 V.
– active in case of error, output signal > 23 mA or > 10,5 V.
.
.
.
.