speed IN
direction
input 1
start
+39VDC
+26VDC
remote
input 2
GND
8 7 6 5 4 3 2 1
15 14 13 12 11 10 9
output 2
speed OUT
output 1
+5VDC
speed intern
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numériques (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogiques
Pin 5, speed IN
0 – 5 V
/ 0 – 10 V
DC
DC
0 – 20 mA / 4 – 20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analoge Ausgänge
Analog outputs
Sorties analogiques
Pin 9, speed OUT
0 – 10 V
/ 0 – 12 kHz
DC
34
Analogschnittstelle
Pin 9, speed OUT
Die werkseitige Einstellung ist
0 –10 V
, proportional zur Motor-
DC
drehzahl 0 – 6000 min
.
–1
Alternativ steht ein Frequenz-
bereich von 0 – 12 kHz zur Ver-
fügung. Wahlmöglichkeit mittels
Schiebeschalter S2 im Gerätein-
nern (siehe Seite 35)
Pin 10, +5 V
DC
Es stehen ca. +5 V
zur Verfü-
DC
gung. (max. Strom 0.5 A)
Pin 13, speed intern
Abhängig von der Betriebsart hat
Pin 13 unterschiedliche Funktio-
nen:
Analogschnittstelle nicht aktiviert
(Normalbetrieb, d.h. Pin 2 offen)
Hier dient Pin 13 als Autostart-
funktion. Ist Pin 13 mit Pin 1
(GND) verbunden, kann die Pum-
pe direkt über die Netzspannung
gestartet bzw. angehalten werden
(Netzschalter muss eingeschaltet
sein).
Analogschnittstelle aktiviert
(Pin 2 auf GND)
- Pin 13 offen:
Die Drehzahl wird über Pin 5
(speed IN) vorgegeben.
- Pin 13 auf GND
Die Drehzahl kann am Bedie-
nungspanel der Pumpe einge-
stellt werden.
Pin 14, output 1 (PWM)
Pin 15, output 2 (PWM)
Digitale Ausgänge in Verbin-
dung mit Pin 7 (+39 V
), als
DC
Pulsweiten-Modulation ausge-
führt (Open Collector). Können
nur über RS232-Schnittstelle be-
dient werden; z.B. mit einer Soft-
ware wie ProgEdit .
Analog interface
Pin 9, speed OUT
The default setting is 0 –10 V
proportionally to the motor speed
0 – 6000 rpm.
Alternatively a frequency range
from 0 – 12 kHz is available. Al-
ternatives with sliding switch S2
inside the pump (see page 35)
Pin 10, +5 V
DC
About +5 V
are available.
DC
(max. current 0.5 A)
Pin 13, speed intern
Depending on how the pump is
operated, pin 13 has different
functions:
Analog interface not activated
(Normal operation, i.e. pin 2 is
open) Pin 13 serves as auto-start
function. If pin 13 is connected
to pin 1 (GND), the pump can be
started and stopped directly from
the power supply (the power
switch must be on).
Analog interface activated
(Pin 2 on GND)
- Pin 13 open:
The rotation speed is adjusted
via pin 5 (speed IN).
- Pin 13 on GND:
The rotation speed can be ad-
justed by the speed selector on
the control panel of the pump.
Pin 14, output 1 (PWM)
Pin 15, output 2 (PWM)
Digital outputs in connection with
pin 7 (+39 V
), implemen-ted as
DC
pulse-width-modulation (open
collector). Can only be activated
via RS232 interface, e.g. with a
software like the ProgEdit.
Interface analogique
Pin 9, speed OUT
,
Le réglage d'usine par défaut est
DC
0 –10 V
, proportionnel au nombre
CC
de tours du moteur 0 – 6000 t/min.
Une zone de fréquence de 0 – 12
kHz est à disposition en alternative.
Possibilité de sélection au moyen
d'un interrupteur coulissant S2 à
l'intérieur de l'appareil (v.p. 35).
Pin 10, +5 V
CC
Environ +5 V
sont à disposition
CC
(courant maximal 0.5 A)
Pin 13, speed intern
Le pin 13 possède des fonctions
différentes en fonction du mode
d'opération:
Interface analogique non activée
(Mode d'opération normal, c.-à-d.
pin 2 ouvert) Le pin 13 a la fonc-
tion „auto-start". Si le pin 13 est
connecté au pin 1 (GND), la pompe
peut être mise en service ou arrêtée
directement par l'alimentation élec-
trique (l'interrupteur principal doit
être sur ON).
Interface analogique activée
(Pin 2 sur GND)
- Pin 13 ouvert: La vitesse de rota
tion doit être ajustée par le pin
5 (speed IN).
- Pin 13 sur GND: La vitesse de
rotation peut être ajustée par le
sélecteur de vitesse sur le tab-
leau de commande de la pompe.
Pin 14, output 1 (MLI)
Pin 15, output 2 (MLI)
Sorties numériques en relation
avec le pin 7 (+39 V
CC
œuvre comme modulation de
largeur d'impulsion (collecteur
ouvert). Ne peuvent être activées
que par le biais de l'interface RS232,
p.e. avec le logiciel ProgEdit.
MCP-Z Process / ISMATEC SA / 03.08.07/CB/GP