Error message
Meaning
The test strip is stuck or the test strip
drum does not rotate.
Press the Test button to turn the meter off.
Open the drum compartment cover and (if
applicable) pull out the test strip. Press the
red drum release button and remove the
test strip drum. Insert the test strip drum
again and close the drum compartment
cover. The meter rotates the drum and
reads the bar code. When you next turn on
the meter to perform a test, the drum
symbol and flashing bottle symbol are
displayed. You may use all the test strips
remaining in the drum. The 90-day use-by
period will no longer be calculated for this
drum. The drum symbol and bottle symbol
will therefore be displayed each time you
perform a test until you insert a new test
strip drum.
192
Symbols, error messages and troubleshooting