6 M enus and settings
6.1 M enu structure of pH/ion (only SG98)
1.
2.
6.2 M enu structure of DO
1.
2.
3.
6.3 T emperature settings
●
Set MTC temperature
If the meter does not detect a temperature probe, MTC appears on the display. In this case the
sam p le temperature should be entered manually. An MTC value between -30 °C and 130 °C
can be entered.
●
Temperature unit
Select the temperature unit: °C or °F. The temperature value is automatically converted
between the two units.
6.4 p H/ion calibration settings (only SG98)
6.4.1 B uffer groups / Standards
6.4.1.1 P redefined pH buffer groups
One of seven predefined buffer groups can be selected:
B1
B2
B3
B4
B5
B6
B7
Temperature tables for these buffers are programmed in the meter and can be found in the "Appen-
dix".
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 27
Temperature Settings
1. Set MTC Temperature
2. Temperature Unit
Calibration Settings
1. Buffer Group/Standards
2. Calibration Mode
3. Calibration Reminder
Temperature Settings
1. Set MTC Temperature
2. Temperature Unit
Calibration Reminder
Measurement Settings
1. Salinity Compensation
1.68
4.01
7.00
2.00
4.01
9.00
2.00
4.00
7.00
1.679
4.008
6.865
1.680
4.008
6.865
1.09
4.65
6.79
1.680
4.003
6.864
3.
Measurement Settings
1. Measurement Resolution
2. Stability Criterion
3. Ion Measurement Unit
4. Rel. mV Offset
4.
Endpoint Formats
5.
Measurement Limits
6.
Sensor ID/SN
2. Barometric Compensation
3. Barometric Unit
4.
Endpoint Formats
5.
Measurement Limits
6.
Sensor ID/SN
10.01
(at 25°C)
9.21
11.00
(at 25°C)
9.00
12.00
(at 20°C)
9.180
(at 25°C)
9.184
12.454 (at 25°C)
9.23
12.75
(at 25°C)
9.182
12.460 (at 25°C)
Mettler US
Mettler Europe
Standard Merck
buffer
JIS Z 8802
DIN19266
DIN19267
Chinese
Menus and settings
29