Interfering Substances Of The Inlab® Optiox Sensor - Mettler Toledo SevenGo pro Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Do not remove the OptiOx cap after installation until a cap replacement is necessary.
8.6  I nterfering substances of the InLab® OptiOx sensor
Following substances may interfere with the dissolved oxygen measurement:
Alcohols greater than 5%,
hydrogen peroxide (H
sodium hypochlorite (NaClO
gaseous sulphur dioxide (SO
Carbone dioxide (CO
), chloride (Cl
ments.
Do not use the InLab
chlo r o f orm or methylene chloride.
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 39
O
) greater than 3%,
2
2
) greater than 3%,
3
) and gaseous chlorine (Cl
2
), ammonia (NH
), pH, any ionic species like sulphide (S
2
3
) and hydrogen sulphide (HS
-
OptiOx sensor in solutions that contain organic solvents, such as acetone,
®
).
2
) do not interfere with the dissolved oxygen mea s ure-
-
), sulphate (SO
2-
2-
4
Maintenance
41
21.04.2011 11:46:18

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sevengo optioxSevengo duo proSevengo duo optiox

Table des Matières