If the "automatic endpoint" format is selected, the measurement stops automatically as soon
as the Stability icon appears.
If the "manual endpoint" format is selected, press READ to manually stop the measurement.
If the "timed endpoint" format is selected, the measurement stops after the preset time.
4.11  T emperature compensation
We recommend the use of either a built-in or a separate temperature probe. If a temperature probe
is used, ATC and the sample temperature are displayed. If no temperature sensor is used, MTC
is dis p layed and the sample temperature should be entered manually. The meter accepts only an
NTC 30 kΩ temperature sensor.
The meter calculates the temperature-adjusted electrode slope using this temperature and shows
the temperature-compensated pH/ion value in the measurement display (only SG98).
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 21
As soon as the measurement is stable according to the selected stability criterion, the Sta-
bil i t y icon appears.
Operating the pH/ORP/Ion/DO meter SG98 and DO meter SG9
23
21.04.2011 11:46:01